Bursting pressure of triple-lumen central venous catheters under static and dynamic loads

Abstract

Introduction

Central venous catheter (CVC) access is commonplace in intensive care units. Patients undergoing computed tomographic angiography require injection of contrast at high flow rates (4.5 mL/s), often CVC access is not used due to safety concerns. The CVC might rupture at high flow rates, resulting in CVC fragmentation and embolization or contrast extravasation.

The objective of this study is to determine the pressure required to burst a CVC under static load and compare this to the pressure generated by injection of contrast at high flow rates (dynamic load) through the distal (16-g) lumen of a triple-lumen CVC.

Methods

We gathered 16-cm long triple-lumen CVCs (n = 14) from patients with an average dwell time of 5.2 days (±2.7 days). Half the CVCs (n = 7) were subjected to static testing, where the distal lumen was occluded with the guidewire and super glue at the distal end of the catheter. The CVC was then placed into a 10-cm deep water bath at 37°C to simulate in vivo conditions and water was injected until catheter rupture. Dynamic pressure testing was done with the remaining catheters, with radio-contrast injected through the unoccluded distal lumen at flow rates of 4.5 mL/s, then 7 mL/s. Pressures were recorded throughout injection.

Results

During static testing, 6/7 CVCs burst at the distal lumen where the glue was applied, the remaining CVC burst proximal to the hub. PSI at burst during static testing was 184.2 PSI (95% confidence interval [CI] 174.3-194.1 PSI). During dynamic testing the mean peak pressures at 7 mL/s was 81.1 PSI (95% CI 73-89.2 PSI). At 4.5 mL/s the mean peak pressure was 47.9 PSI (95% CI 42.9-52.9 PSI).

Conclusions

No CVCs failed under dynamic loading with injection of contrast at flow rates (4.5 and 7 mL/s) high enough to support computed tomographic angiography. This suggests 16-cm triple-lumen CVCs can be used safely.

Post author correction

Article Type: ORIGINAL RESEARCH ARTICLE

Article Subject: Intensive care

DOI:10.5301/jva.5000776

Authors

Tim Beckingham, Andy Roberts, Alana St. John, Gerry O’Callaghan

Article History

Disclosures

Financial support: None.
Conflict of interest: None.

This article is available as full text PDF.

  • If you are a Subscriber, please log in now.

  • Article price: Eur 36,00
  • You will be granted access to the article for 72 hours and you will be able to download any format (PDF or ePUB). The article will be available in your login area under "My PayPerView". You will need to register a new account (unless you already own an account with this journal), and you will be guided through our online shop. Online purchases are paid by Credit Card through PayPal.
  • If you are not a Subscriber you may:
  • Subscribe to this journal
  • Unlimited access to all our archives, 24 hour a day, every day of the week.

Authors

Affiliations

  • Lyell McEwin Intensive Care Unit, Adelaide, South Australia - Australia

Article usage statistics

The blue line displays unique views in the time frame indicated.
The yellow line displays unique downloads.
Views and downloads are counted only once per session.

No supplementary material is available for this article.